I. PURPOSE
To describe the procedure
for washing of HPLC column.
II. SCOPE
This procedure is applicable to HPLC columns in Quality Control department,
III. RESPONSIBILITY
Assistant, Executive - Quality Control
IV. ACCOUNTABILITY
Incharge - Quality Control
V.
PROCEDURE
1.0
Instructions
1.1
After analysis, wash the column completely salt free.
1.2
Before storing, the column should be washed with 100%
organic solvent (Acetonitrile/
Methanol).
2.0 FOR
C18, C8 and PHENYL COLUMNS
2.1 After usage , column
should be washed with Milli-Q water at the flow rate of 1 ml/min for about 60 min.
2.2
After water washing, wash with 100% solvent
(Acetonitrile / Methanol) at the flow rate of 1 ml/min for minimum 30 min.
2.3
Check that the pressure does not exceed 3000 PSI.
2.4
Enter the column usage details in ‘HPLC column usage
log’
Format No.:
F01/SOPQC034-00.
2.5
Fix the end fittings properly and store in an
appropriate place.
2.6
Before every use, wash the column with milli-Q water at
the flow rate of 1ml/min for about 30 min.
3.0 List
of Formats
Format No.
|
Title
|
F01/SOPQC034-00
|
HPLC Column Usage
Log
|
VI. REVISION
SUMMARY
Version No.
|
Revised on |
Reason for Revision |
00
|
NA
|
NA |