I.
PURPOSE
To describe a procedure for
calibration of QC instruments.
II.
SCOPE
This procedure is applicable for all the instruments
which required periodic calibration in
Quality Control department,
III.
RESPONSIBILITY
Assistant, Executive- Quality Control
IV.
ACCOUNTABILITY
Incharge –Quality Control
V.
PROCEDURE
1.0
Calibration falls in two categories
1.1
Internal calibration : which is carried out by user /
maintenance department of with in house procedures.
1.2
External calibration : which is carried out by third
party.
2.0
Annual calibration schedule should be prepared as
per
Annexure No. AX01/SOPPQC006-00.
3.0
Before starting the calibration, check for the
availability and validity of the required standards/equipments/instruments.
4.0
Ensure that all
the standards / instruments/equipments are traceable to national /
international standards (where applicable).
5.0
Before starting
the calibration, go through the calibration procedure and protocol.
5.1
Perform the calibration as per ‘Calibration procedure’.
5.2
The instruments which are covered under daily
calibration should be labeled with
“DAILY CALIBRATION” label and the calibration details should be
entered in the respective standard format.
6.0
After completion of Calibration/Validation,
6.1
If the results found satisfactory, label the instrument
with calibration status for calibration and validation status label for
validation.
6.2
Write down all the observations / results in protocol
and take approval of authorized person.
6.3
Before approval of records, authorized person should
check all the calculation and ensure that the calibration is carried out as per
calibration procedure and the results obtained are well within specified
limits.
6.4
If the results are not satisfactory, label the
instrument with
“OUT OF CALIBRATION” label, and for
validation,
“OUT OF VALIDATION” label.
6.5
If the results are not satisfactory, inform to incharge
Q.C.
6.6
Investigation should be carried out as per ‘Out Of
Specification’ (OOS) SOP.
6.7
Action plan
to be followed by incharge QC
6.7.1
Ensure that standard procedure is followed for
calibration/validation.
6.7.2
Check whether all the calculations are correct. (if
applicable).
6.7.3
Ensure that right standards/equipments/instruments are
used for
calibration/validation.
6.7.4
Enter all the data in tabular format for evaluation.
6.7.5
Head -QA should write conclusion after reviewing the
data.
6.8
Satisfactory
6.8.1
Calibration / validation results are within specified
limits as mentioned procedures.
6.9
Not
satisfactory
6.9.1
Calibration / validation results are not within
specified limits
as
mentioned in procedures.
7.0
List of Annexures
Annexure
No.
|
Title |
AX01/SOPQC 006-00
|
Annual calibration
schedule
|
VI.
REVISION SUMMARY
Version No.
|
Revised on
|
Reason for Revision |
00
|
NA
|
NA
|