I PURPOSE
To describe the procedure for cleaning and
maintenance of Laminar air flow.
II SCOPE
This
procedure is applicable to Quality Control department-Penicillins.
III. RESPONSIBILITY
Microbiologist – Quality Control
IV. ACCOUNTABILITY
Incharge - Quality Control
V PROCEDURE
1.0
Cleaning of laminar air flow (LAF) -daily
2.0 Clean the UV lights, tube light and
manometer with sterile mop.
3.0 Clean the working table, top and side
glass with filtered 70% isopropyl alcohol on the grill of laminar
flow.
4.0 After cleaning, keep the UV lights “ON”
when laminar flow is not in operation.
5.0 Manometer reading is between 8.0 and 14.0
mm .in ‘ON’ condition and zero in ‘OFF’ condition.
6.0 Maintain a register F01 to record positive
pressures, contents are as per Table-1.
7.0 Record the cleaning details in register
F02 and the contents are as per Table-2.
8.0 If manometers reading are out of the
limit, inform maintenance department for cleaning of filters.
Table-1(F01/SOP PMB 004)
Title: MANOMETER USAGE REGISTER
COLUMN
No.
|
NAME OF THE COLUMN |
SHORT FORM |
COLUMN
WIDTH(cm)
|
|||||
LAF-701 |
||||||||
1 |
----
|
1.5
|
||||||
2
|
DATE
|
----
|
1.5
|
|||||
3
|
SWITCHED ON/OFF
|
FROM
|
----
|
1.5
|
||||
TO
|
1.5
|
|||||||
4
|
MANOMETER RANGE
|
----
|
3.0
|
|||||
LAF-702 |
||||||||
1
|
SWITCHED ON/OFF
|
FROM
|
----
|
1.5
|
||||
TO
|
1.5
|
|||||||
2
|
MANOMETER RANGE
|
----
|
3.0
|
|||||
3
|
CHECKED BY
|
----
|
2.0
|
|||||
4
|
REMARKS
|
----
|
LEFT OVER SPACE
|
|||||
Table-2 (F02/SOP PMB
004)
Title: LAF CLEANING REGISTER
COLUMN
No.
|
NAME OF THE COLUMN |
SHORT FORM |
COLUMN
WIDTH(cm)
|
||
LAF –701
|
|||||
1 |
----
|
1.5
|
|||
2
|
CLEANED ON
|
----
|
1.5
|
||
3
|
CLEANED BY
|
----
|
3.0
|
||
LAF –702
|
|||||
1
|
CLEANED ON
|
----
|
1.5
|
||
2
|
CLEANED BY
|
----
|
3.0
|
||
3
|
CHECKED BY
|
----
|
3.0
|
||
4
|
REMARKS
|
----
|
LEFT OVER SPACE
|
||
VI LIST OF FORMATS
FORMAT
NUMBER
|
PURPOSE
|
F01/SOP PMB 004
|
MANOMETER USAGE REGISTER
|
F02/SOP PMB 004
|
LAF CLEANING REGISTER
|
VII REASON FOR REVIEW
Periodic
review
VIII REVISION
SUMMARY
Version No.
|
Revised on |
Reason for Revision |
02
|
NA
|
NA |
It is really a nice and useful piece of info. I'm happy that you shared this useful info with us. Please stay us up to date like this.
ReplyDeleteThanks for sharing!
Acmas Technologies Inc is leading laminar airflow cabinet or vertical autoclave manufacturer in India.
This is the most useful post! Keep Continue like this post. Thanks for Sharing.
ReplyDeleteoutpatient clinic room ventilation | Ultra-clean air system operating room