I. PURPOSE
To define
the Standard Operating Procedure for Cleaning of Sparkler Filter.
II. SCOPE
This
procedure is applicable for Sparkler filter used in Production Department at Unit-V.
III. RESPONSIBILITY
Assistant ,
Executive - Production
IV. ACCOUNTABILITY
Head -
Production
V.
PROCEDURE
1.0 Preparation of buffers
1.1 pH
7.0
1.1.1
Dissolve one tablet / Capsule of pH 7.0 buffer in
100ml of distilled water and see the pH at 20°C.
1.1.2
The pH should be 7.0.
1.2 pH 9.2
1.2.1
Dissolve one tablet / Capsule of pH 9.2 buffer in
100ml of distilled water and see the pH at 20°C.
1.2.2
The pH should be 9.2.
1.3 pH 4.0
1.3.1
Dissolve one tablet / Capsule of pH 4.0 buffer in
100ml of distilled water and see the pH at 20°C.
1.3.2
The pH should be 4.0.
1.4 Prepared buffers should be given use
before date for one week, if any growth observed before use before date, discard the solution and prepare new
solution.
2.0 Operation of
pH meter
2.1
Ensure that the working area and the instrument is
clean.
2.2
Check the calibration status of the instrument.
2.3
Switch on the instrument.
2.4
Ensure the electrode is properly connected.
2.5
Clean the electrode thoroughly with purified water and
wipeout using tissue paper.
2.6
This cleaning procedure should be followed at each
time of measurement.
2.7
Take the sample solution in a clean dry beaker.
2.8
Set the temperature as per sample specification.
2.9
Mix the sample thoroughly and place the electrode in
the sample solution. The pH will be displayed and note the pH value.
2.10
After measuring the pH clean the electrode with
purified water.
2.11
Always keep the electrode in distilled water.
2.12
Switch off the instrument when it is not in use.
2.13
Electrode
Maintenance
2.13.1
Inspect the electrode frequently for scratches,
membrane coating, and salt crystal buildup.
2.13.2
Rinse off salt deposits or membrane deposits with
purified water.
2.13.3
If rinsing with purified water does not remove the
deposits, then soak the electrode in 0.01M HCl or 0.1M HNO3 for 30
minutes and finally rinse with purified water.
2.13.4
If the reference electrolyte appears to be
contaminated with extraneous matter drain the reference electrode chamber.
2.13.5
Press the CAL/ PRT Button.
2.13.6
Instrument displays ‘CAL 2’ , press YES.
2.13.7
Now it displays pH Cal 2, Buffer 1 RDY, press YES.
2.13.8
Keep Buffer 7.0 and press YES.
2.13.9
Now it displays buffer 2 RDY.
2.13.10Remove the solution and
clean the Electrode with distilled water and wipe with tissue.
1.1 Keep Buffer 4.0 and press YES.
1.2 System displays the pH value of 4.0
Acceptance limit is 4.0 ± 0.02
1.3 Keep the 9.2 buffer after cleaning the electrode.
2.0 Interpretation of Results
2.1 If the
calibration results are found satisfactory, then the instrument can be used for testing.
2.2 If the results are found out of limits,
adjust the slope of the instrument and recalibrate.
2.3 If the problem still persists follow the
under given procedure. Investigate as per OOS.
2.4 After OOS investigation, if the
calibration results are not satisfactory affix ‘Out
of
Calibration’ Label on the instrument.
2.5 Call the service engineers to rectify
the instrument.
2.6 After rectification, calibrate the
instrument.
3.0 Frequency
At the beginning of each shift
VI.
REVISION SUMMARY
Version No.
|
Revised on
|
Reason for Revision |
00
|
NA
|
NA |
END OF THE
DOCUMENT
I agree with your all information about the ph testing drops and i just appreciate it.
ReplyDeleteph testing drops
Hello Dear..
ReplyDeleteI appreciate your Informative post and It's very helpful.Thanks for sharing Keep it up!
PH Meter
Salt Spray Chamber
Bottle Cutting Tester Machine
Bursting Strength Tester Digital
Cobb Sizing Tester
Dart Impact Tester
Digital Gloss Meter
Drop Tester
Edge Crush Tester Digital
Laboratory Testing Instruments Equipment
Paint Plating Testing Instruments
Plastic Bottle Color Measurement
Puncture Resistance Tester
Roofing Color Measurement
Stiffness Tester
Tearing Strength Tester
Tensile Strength Tester Machine Manufacturer
Top Load Tester
Torque Tester
Vibration Table