SOP for operation of Tap density tester



I.                   PURPOSE

To describe the procedure for operation of Tap density tester.

II.        SCOPE

This procedure is applicable to Tap density tester in Quality Control department,….

III.       RESPONSIBILITY

Assistant, Executive - Quality Control

IV.       ACCOUNTABILITY
Incharge  - Quality Control
V.                PROCEDURE
1.0        System Set Up
1.1       Place the instrument on a leveled study base.
1.2       Place the Cylinder holders on the Anvil of the desired method on ETD-1020.
1.3        Plug the mains cord in the socket provided at the rear side of the instrument.          and the other side to 220/230 V AC, 50/60 Hz power supply socket.
2.0       System Operation
2.1       After the System Set Up is over, switch on the power.
2.2       The Instrument will initialize itself. The display will Flash & show the start up screen.
ELECTROLAB ETD-1020
TAP DENSITY TESTER
2.3       The system is now ready for setting of the test parameter & to run the test.             Using the Digit Scroll Key, the user can set the mode of operation as USP or USER.
MODE=USP/USER                                     
SET = menu, START = run
2.4        In USP mode, the instrument will stop after every programmed Tap Count  value and the validation report is available at the end of the test.
2.4.1        In USER mode, the instrument will stop after the set tap value. No validation report and the result screen is available in this case. Press the SET key to set the test parameter.
2.4.2        Press the START key to run the test.
2.5        Setting the Method:
2.5.1        The Method for the test can be set as USP I or USP II, using the METHOD key on the front panel. This is the Toggle key.  When pressed the display will show the changed Test Method. METHOD indicator on the front panel indicates the selected method for the test.
METHOD = USP I

METHOD = USP II

2.5.2        The METHOD key is active during the SET Menu and when the test is not ON.


2.6            Set Menu- USP Mode: (Tap Count Settings)
2.6.1           The first parameter is for setting Tap Count1. The value of TAP COUNT1 can be set from 1 to 9999 taps using the UP/ DOWN & DIGIT SCROLL key. Press the ENTER key to go to the next parameter.
TAP COUNT 1 : 0000         utw

SET = Exit     /    Enter = Scroll

                                    USP specified 500 Taps as Tap count 1.
2.6.2           The next parameter is for setting the Tap Count2.  The value of TAP COUNT2 can be set from 1 to 9999 taps using the up/ down 7 digit scroll key.  Press the ENTER key to go the next parameter.
TAP COUNT 2 : 0000         utw

SET = Exit      /     Enter = Scroll

                                    USP specified 750 Taps as Tap Count 2
2.6.3           The next parameter is for setting Taps Count3. The value of TAP COUNT3 can be set from 1 to 9999 taps using the UP/DOWN & DIGIT SCROLL key. Press the ENTER key to go the next parameter.

TAP COUNT 3 : 0000         utw
SET = Exit    /    Enter = Scroll
                                                            USP specified 1250 Taps as Tap Count 3
Note :           The User can exit from the SET Menu at any point by pressing the SET key.
2.7              Date & Clock Setting:
2.7.1           The next menu is for Clock setting. The default setting for this menu should be the current Time.  The Time can be adjusted using the UP/DOWN & DIGIT SCROLL key. Press the ENTER key to go to the next parameter.
CLOCK: 00.00 .00           utw
SET = Exit     /       Enter = Scroll

2.7.2              The next menu is for Date setting.  The default setting for this menu should be the current Date. The Date can be changed using the UP/DOWN & DIGITAL SCROLL key.  Press the ENTER key to go to the next parameter.
DATE: 10  Jan  2005     utw
SET = Exit    /     Enter = Scroll

2.7.3           If the Date is invalid, then the instrument will prompt the message as “Invalid Date” for about 2 second and the instrument will restore the last setting the date.

Example of valid Date :          29 FEB 1900
                                                                                    31 APR.1999
Invalid Date

2.7.4              The next menu is for the selection of the opting to Print the Last Test result. If this option is set to “Y”, & ENTER key is pressed then the result of the last test will be printed before the new test has started.
PRINT LAST RESULT? Y/N          w

SET = EXIT    /     ENTER = SCROLL


2.7.5        If the option is set to “Y” & the printer is not connected or ready, then the Instrument will prompt the message every 5 Sec’s till the Printer is made ready or the print option is disabled.  (Set to N & ENTER key is pressed.)
PRINTER NOT RESPONDING     
WANT TO PRINT? Y/N        w


2.8              Exiting from SET menu:
2.8.1              Press the ENTER key to scroll back to SET Menu.  This will lead to point (a) of SET MENU), Press the SET key to exit from the set Menu.
MODE = USP   /    USER                  w
SET = menu ,              START = run
2.9            Set Menu on User Mode: (Tap Count Settings)
2.9.1                 In the USER Mode, the user can set the required number of Taps.  The Taps are programmable from 1 to 9999 taps. The test will be ON till the number of programmed taps.  No test report is available in this case even if the printer is connected.
SET TAPS =    0000                 utw

Press START to run test

2.9.2                In User Mode, Press the START key from the front panel to Start the test.  The display will now show the elapsed Tap. The test will continue till the programmed tap value.  No test printout is available in this case even if the Printer is connected or Online.
ELAPSED TAPS: 0000

Drops / Min . = 000


2.9.3        The test can be halted when in  Run by pressing the STOP key from the front panel.  The test can be continued from the same point by pressing the START key from the front panel.  Pressing the STOP key will abort the test and the Instrument will initialized itself for the new test.

TEST HALTED

START=CONT.   /    STOP = ABORT

2.9.4        Pressing the Start key would start the test as per the method selected. The display will show elapsed Tap and Drops per minute.
ELAPSED TAP : 0000

Drops / Min. = 000

2.9.5      If the Drops per minute deviate by ±10 Counts then an error screen will be displayed. The instrument will halt.  This is possible if there is some problem with cylinder holder movement. Press the START5 key to continue or STOP key to abort the test.

Drops / Min. not Accurate

START = cont.   / STOP = Abort

2.9.6     After the programmed Tap Count 1 is over, measure the tapped volume Va to its nearest graduated unit.  Enter the value for the Va using the UP/DOWN & DIGIT SCROLL keys. Press the ENTER key to register the value of Va.
MEASURE TAPPED VOLUME
VOLUME (Va.) = 0000  ml   utw
2.9.7        Press the Start key to start the test again.  The test will continue till the programmed Tap Count 2 is over.  The display will now show the elapsed Tap and Drops per minute.
PRESS START TO
CONTINUE THE TEST
2.10                 Start Menu-USP Mode:
2.10.1             Press the  START key to run the test. Weigh the test sample and enter the Weight of the test sample.  Press the ENTER key to register the weight.

WEIGH SAMPLE,    ENTER WGT

WEIGHT = 000. 00 gm         utw
2.10.2                Fill the test sample into measuring cylinder. Carefully level the test sample without compacting.  Read the unsettled apparent volume Vo to the nearest graduated unit and enter the volume using UP/DOWN and DIGIT SCROLL key. Press the ENTER key to register the volume.

FILL SAMPLE IN CYLINDER
VOLUME (Vo) = 0000  ml  utw

2.10.3               
KEEP CYLINDER IN HOLDER
 PRESS START TO RUN TEST
 
Keep the measuring cylinder in the holder. Lock the holder assembly. Put the Cylinder Holder on the ETD-1020.Press the Start key to run the test.



2.10.4             If the START key is pressed; and the cylinder Holder is not on the ETD-1020, then an error screen will display.  The instrument will halt. Place the cylinder Holder on ETD-1020 properly & press the START key to continue or STOP key to abort the test.

HOLDER MISSING ,   PLACE IT

START = Cont .   /     STOP = Abort


2.10.5             After the programmed Tap Count 2 is over, measure the tapped volume Vb to its nearest graduated unit. Enter the value for the Vb using UP/ DOWN & DIGITAL SCROLL key.  Press the ENTER key to register the value of Vb.

MEASURE TAPPED VOLUME
VOLUME(Vb) = 0000 ml        utw

2.10.6             If Volume Vb is more than Volume Va, then the instrument will prompt an error message. The operator can select the option either to Abort the test or to reenter the volume.
VOLUME (Va) < VOLUME (Vb)
Test Abort / Reenter?                       w

2.10.7            In case of re enter, the option (f) is displayed again.  This will be continuing till the correct volume is entered OR the test is aborted.

MEASURE TAPPED VOLUME
VOLUME(Vb) = 0000 ml     utw

2.10.8        The difference between the two volumes Va and Vb is now calculated and displayed.  Press the START key to continue the test. Pressing the Stop key will display the results.
Diff (Va- Vb ) = 000. 00 %
START = CONT. / STOP = ABORT
Note:   USP has specified to run the test till the difference between the two successive volume measurements (Va & Vb) is less than 2%.
2.10.9        Pressing the START key will continue the test till the programmed Tap Count 3 is over.  The display  will now show the elapsed Tap and Drops per minute. In this case the last entered volume Vb becomes Va.                                   

ELAPSED TAP : = 0000

Drops / Min . = 000


2.10.10  After the programmed Tap Count 3 is over, measure the  tapped   
              volume Vb to its nearest graduated unit.  Enter the value for the      Vb using the UP / DOWN & DIGITAL SCROLL keys. Press the        ENTER key to register the value of Vb.
MEASURE TAPPED VOLUME
VOLUME (Vb) =  0000 ml    utw

2.10.11 The difference between two volume & Vb is calculated again and  
             displayed.  Pressing the START key will continue the test again                  for the Tap Count 3 value.  Pressing the Stop key will display the            results.

Diff (Va- Vb ) = 00.000 %
START = CONT. / STOP = ABORT

2.10.12 Pressing the STOP key the results like Tapped Density  
             (T.Density), initial Density (Init. Den.)  Hausner Ratio,       Compressibility Index etc. are calculated and displayed.
T. Density = = 0.0000 gm/ml
Set = New Test ,                u = I.D.
2.10.13 Pressing the DOWN key, the result of different parameters will be displayed one after the other. The Validation report of the complete test is available if the Printer is connected and online.

Comp. Index  =  000.00 %
Set = New Test,                  u = H.R.


Hausner Ratio  =  00. 000
Set = New Test ,                u = T.D.
2.10.14  If the printer is not connected or not Ready, then the                 
             Instrument will prompt the message every 5 Sec till the printer is     made ready or the print option is disabled (Set to “N” & the           ENTER key is pressed)
PRINTER NOT RESPONDING
WANT TO PRINT ? Y/N            u
2.10.15 Pressing the Set key from the front panel will initialize the instrument for the new test.
MODE=USP/USER                       u
SET = menu ,        START = run. 
2.11        Precautions
2.11.1                Avoid any dust gathering on the bearings.
2.11.2                Do not lubricate the bearings.
2.11.3                Keep the shaft and the tip of the cylinder holder clean.
2.11.4                Do not press the cylinder holder when the tapping base of cylinder holder is not touching the anvil.
2.11.5                Place the cylinder in the holder only when the holder is placed on the anvil.
 VI.      REVISION SUMMARY
Version No.

Revised on

Reason for Revision

00
NA

NA

 

END OF THE DOCUMENT